Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Amprobe RS-3: (1 year) CAL VER /5500 DATE: 25-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 10 NUMBER OF LINES: 95 CONFIGURATION: Fluke 5500A STANDARD: Fluke 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # rs3v.bmp # rs3z.bmp # rs310.bmp # rs350.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- P F V 1.003 ASK+ R K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Insert bayonet type voltage test leads into Voltage 1.007 DISP receptacles at bottom of unit. Push against receptacle 1.007 DISP spring and twist to lock in place. The right receptacle 1.007 DISP is positive, the left receptacle is negative. 1.008 HEAD {AC VOLTAGE} 1.009 DISP ADJUST the front panel zero adjust screw for a meter 1.009 DISP indication of exactly zero. 1.010 PIC rs3v 1.011 HEAD AC VOLTAGE: {150V Range} 1.012 5500 150 150V 3/ 60H SI 2W 2.001 HEAD AC VOLTAGE: {300V Range} 2.002 5500 300 300V 3/ 60H SI 2W 3.001 HEAD AC VOLTAGE: {600V Range} 3.002 5500 600 600V 3/ 60H SI 2W 4.001 HEAD {RESISTANCE} 4.002 DISP Disconnect the test lead from the right voltage 4.002 DISP receptacle at the bottom of the instrument. 4.002 DISP 4.002 DISP Insert the Ohmmeter battery/fuse attachment to the 4.002 DISP jack on the right side of the instrument, just below 4.002 DISP the Ohmmeter zero knob. 4.003 RNG 150V AC 4.004 PIC rs3z 4.005 DISP Check the meter zero for an INFINITE indication. 4.005 DISP 4.005 DISP If necessary adjust meter zero screw. 4.006 ASK- U #2 wire comp. uncecessary since lead resistance is zeroed at uut. 4.007 5500 0.0Z S 2W 4.008 DISP Adjust the OHMS ZERO ADJUST KNOB for a "0" indication. 4.008 DISP 4.008 DISP If Ohmmeter will not zero, check fuse and battery 4.008 DISP in the Ohmmeter attachment. 4.009 5500 100Z 5% 2W 5.001 ASK+ U 5.002 HEAD {AC CURRENT} 5.003 DISP Loop a piece of heavy gauge copper wire from the 5500A 5.003 DISP AUX HI to the AUX LO terminal. Clamp the meter around 5.003 DISP the wire aligning the wire with the marks on the clamp. 5.004 HEAD AC CURRENT: {6A Range} 5.005 5500 6 6.00A 3/ 60H SI 2W 6.001 HEAD AC CURRENT: {15A Range} 6.002 PIC rs350 6.003 5500 15 15.0A 3/ 60H SI T5 7.001 HEAD AC CURRENT: {40A Range} 7.002 5500 40 40.0A 3/ 60H SI T5 8.001 HEAD AC CURRENT: {100A Range} 8.002 5500 100 100A 3/ 60H SI T5 9.001 HEAD AC CURRENT: {300A Range} 9.002 5500 300 300A 3/ 60H SI T5 10.001 END